Application |
Sample type |
Input recommended |
Whole genome sequencing |
Plant or animal gDNA (without serious degradation) |
50-500 ng |
Cell-free DNA |
≥1 ng (analytical sensitivity dependent) |
|
Microbial DNA (without serious degradation) |
1-500 ng |
|
Targeted sequencing |
gDNA (without serious degradation) |
≥ 10 ng (quality dependent)* |
FFPE DNA (with average size>500 bp) |
≥ 100 ng (quality dependent)* |
|
Cell-free DNA |
≥ 1 ng (analytical sensitivity dependent) |
Workflow
High Efficiency
Fig 1. Library yield with different types of adapters. The yields of the libraries prepared with different gDNA initial amount by using NadPrep DNA Library Preparation Kit (for Illumina®) coupled with NadPrep Universal Stubby Adapter (UDI) Module or NadPrep UDI Adapter Module (for Illumina®) Note: The sample is human female genomic DNA (Promega-Female, G1521).
Unbiased Coverage
Fig 2. Uniform coverage of regions with various GC content. Coverage distribution for libraries generated from 10 ng DNA of A. Staphylococcus aureus (GC%=33), B. Salmonella enterica LT2 (GC%=52) and C. Pseudomonas aeruginosa (GC%=67). GC% distribution per 100 bp bin is indicated by histograms and normalized coverage is represented by lines.
Accurate Variant Analysis
Fig 3. Variant calling in reference standard DNA. The DNA libraries were prepared from Quantitative Multiplex Formalin Compromised (Mild) DNA (Horizon Discovery, HD798). The enriched libraries were sequenced on HiSeq X Ten (PE150) and subsampled to 5 Gb for analysis. Vardict was used to call variants.
NadPrep DNA Library Preparation Module (for Illumina)
/
/
Package
Color of Tube Cap
Component
Volume
1002101 (24 rxn)
Volume
1002103 (96 rxn)
Box1
EndRepair & A-Tailing Buffer
180 µL
690 µL
End Repair & A-Tailing Enzyme
120 µL
460 µL
Ligation Buffer
750 µL
2×1500 µL
DNA Ligase
65 µL
250 µL
2X HiFi PCR Master Mix
720 µL
2×1600 µL
NadPrep
Amplification Primer Mix
145 µL
650 µL
Nuclease Free Water
2.5 mL
8 mL
TE Solution
1 mL
4 mL
Box2
NadPrep SP Beads
4 mL
15 mL
NadPrep UDI Adapter Module
Color Of Tube Cap | Component |
Volume 1003211 (24 rxn) |
Volume 1003222 (24 rxn) |
Volume 1003211(96 rxn) |
Volume 1003212(96 rxn) |
Volume 1003213(192 rxn) |
Volume 1003224(480 rxn) |
|
NadPrep UDI Adapter # | 12×10 µL | 12×10 µL | 12×22 µL | 12×22 µL | 24×22 µL | 96×3.5 µL (Plate)×5 |
Universal Stubby Adapter (UDI) Module
Color Of Tube Cap | Component |
Volume 1003240 (24 rxn) |
Volume 1003241 (96 rxn) |
Volume 1003242 (96 rxn) |
|
NadPrep Universal UDI-index Primer Mix # | 12×15 µL | 24×25 µL | 24×25 µL |
|
NadPrep Universal Stubby Adapter | 60 µL | 230 µL | 230 µL |
|
NadPrep Amplification Primer Mix II | 20 µL | 75 µL | 75 µL |
Color of Tube Cap | Component | Volume | Volume | Volume | Volume |
1003251 (96 rxn) | 1003252 (96 rxn) | 1003253 (96 rxn) | 1003254 (96 rxn) | ||
|
NadPrep Universal UDI-index Primer Mix # | 96×8 µL (Plate) | 96×8 µL (Plate) | 96×8 µL (Plate) | 96×8 µL (Plate) |
|
NadPrep Universal Stubby Adapter | 260 µL (Tube) | 260 µL (Tube) | 260 µL (Tube) | 260 µL (Tube) |
|
NadPrep Amplification Primer Mix II | 75 µL (Tube) | 75 µL (Tube) | 75 µL (Tube) | 75 µL (Tube) |
Possible Reason |
Solution |
The DNA sample contains more than 1 mM of EDTA |
Purify the DNA by using magnetic beads or spin columns,and dissolve the purified DNA with TE Solution. |
The DNA sample contains impurities |
High quality DNA input, with OD260 /OD280 =1.8-2.0 and OD260 /OD230 =2.0-2.5. The residual guanidine salt and proteins may suppress the fragmentase activity. Purify the DNA by using magnetic beads or spin columns,and dissolve the purified DNA with TE Solution. |
Fail to thoroughly mix the reaction mix |
Mix thoroughly and centrifuge to collect the contents to the bottom of the tube before initiate the reaction. |
Fragments self-ligation |
Thoroughly read the manual before operation. |
Possible Reason |
Solution |
The DNA is seriously degraded |
Degraded DNA, such as FFPE C DNA, may result in short fragments (150-200 bp) without change of the same fragmentation condition. |
The fragmentation process lasts too long |
Avoid operating multiple samples at one time. |
The operation under room temperature lasts too long |
Ensure all the processes operated on ice beforeloading into the thermo cycler. |
Possible Reason:Inappropriate dilution of adapters
Solution:For low DNA input, adapters should be diluted as recommended prior to use.
Type
Product
Scale
Catalog
Library Preparation Module
NadPrep DNA Library Preparation Kit (for Illumina®️) G24
24 rxn
1002101
NadPrep DNA Library Preparation Kit (for Illumina®️) E96
96 rxn
1002103
Universal Stubby Adapter (UDI) Module
NadPrep Universal Stubby Adapter (UDI) Module Set A1, 24 rxn
Index 1-12,
24 rxn
1003240
NadPrep Universal Stubby Adapter (UDI) Module Set B1, 96 rxn
Index 1-24,
96 rxn
1003241
NadPrep Universal Stubby Adapter (UDI) Module Set B2, 96 rxn
Index 25-48,
96 rxn
1003242
NadPrep Universal Stubby Adapter (UDI) Module Set E1, 96 rxn
Index 1-96,
96 rxn
1003251
NadPrep Universal Stubby Adapter (UDI) Module Set E2, 96 rxn
Index 97-192,
96 rxn
1003252
NadPrep Universal Stubby Adapter (UDI) Module Set E3, 96 rxn
Index 193-288,
96 rxn
1003253
NadPrep Universal Stubby Adapter (UDI) Module Set E4, 96 rxn
Index 289-384,
96 rxn
1003254
UDI Adapter Module
NadPrep UDI Adapter Kit, Set A (for Illumina®️)
Index 1-12,
300 pmol / Adapter
1003211
NadPrep UDI Adapter Kit, Set B (for Illumina®️)
Index 13-24,
300 pmol / Adapter
1003212
NadPrep UDI Adapter Kit, Set A&B (for Illumina®️)
Index 1-24,
300 pmol / Adapter
1003213
NadPrep UDI Adapter Kit Set C1 (for Illumina®️)
24 rxn
1003221
NadPrep UDI Adapter Kit Set D1 (for Illumina®️)
480 rxn
1003224
Nanodigmbio will contact you within 1 day.